Description of the method:
The modified Couette measurement principle is used. The method has a density cell and a dynamic viscosity cell. The kinematic viscosity is calculated automatically, based on the dynamic viscosity and the density of the sample.
Interpretation of the results:
Viscosity is usually determined at 40 and 100 ° C, on this basis the viscosity index can be determined – a parameter describing the dependence of viscosity on temperature. WL is characteristic of various types of oils. The oil base and the type of additives used are important. The viscosity test tells the user whether the oil used is in compliance with the requirements. The increase in viscosity can indicate contamination / mixing, aging, evaporation of the oil, decrease in viscosity and in turn of dilution / mixing, cracking or mechanical degradation (shear stress).