Description of the method:
A previously prepared sample in the consistency profile test using the GTA apparatus – the lubricant film on the substrate created in die extrusion – is analyzed using a inteligent vision system. Thanks to the advanced recording system and developed computational algorithms, the degree of lubricant contamination is classified by determining the particle concentration (in mg/g in sizes >10 µm, >25 µm, >80 µm) and the percentage of particles with sizes >25 µm and the classification of their formation processes ( e.g. abrasive wear, adhesive wear, fiber wear).
Interpretation of the results:
The test results illustrate the weight distribution of particles in individual distribution compartments (>10 µm, >25 µm, >80 µm), and characterize particles with a size >25 µm, i.e. at a detection level that cannot be determined by elemental emission spectroscopy (RDE). -OES). Based on the results, conclusions can be drawn about the occurrence of external contamination and the level of advancement of wear processes in the lubricated connection, especially when observing the trend.